Product description
- Key valve test functions include pre-configured valve signature test, speed test, step test, manual test and bump/partial stroke test
- Key mA loop calibration functions include mA generate, mA simulate, mA read, mA read/loop power and voltage read.
- The ValveTrack software allows data to be uploaded to a PC for further analysis of valve measurements logged and recorded in the 710's memory
Testing valves easier than ever
The Fluke 710 loop calibrator for valve testing is designed for quick and easy testing of control valves with HART technology. The 710's integrated test procedures and intuitive user interface allow users to test valves quickly and easily. In addition, the quick test results provide a clear diagnosis for quick maintenance decisions. Based on the results of the quick valve test, you know whether your valve is in good, marginal or poor condition, so you can quickly decide whether additional maintenance is required.
Valve testing and HART communication with a precision loop calibrator
The integrated HART communication function of the 710 loop calibrator for valve testing allows users to generate a 4-20 mA signal that controls the smart control valve. At the same time, the HART feedback signal is interpreted to determine whether the valve moves to the expected position. In addition to valve position information, the HART communication protocol can be used to determine the pressure delivered by the valve that moves the valve.
The 710 provides integrated test procedures that automatically increase and change the mA signal while monitoring HART feedback on the valve position and pressure. This gives you a better picture of the valve's condition at the simple touch of a button.
Preconfigured valve tests, concise answers
Valve testing functions integrated into the 710 include
- Manual testing; manual modification of the mA signal and information on the HART position and pressure variable
- Increase of the mA signal over the full range from 4 to 20 to 4 mA during registration of the 0-100-0% position or of the pressures that move the valve from 0-100-0%
- Step change of the mA signal at the input of the valve and evaluation of the response of the valve to changes in the mA input signal
- Velocity tests to determine how fast the valve can be opened and closed
- Impact and partial stroke tests that help test valves over part of their range so they can be tested during an actual process
ValveTrack software allows further analysis and trends to be created
Valve tests logged and stored in the 710's memory can be uploaded using the ValveTrack analysis software.The ValveTrack software provides you with the following capabilities
- Uploading, printing and plotting of logged on-site valve tests performed
- Comparison of previously uploaded tests with recent tests
- Display of valve test history via HART label ID
- Export valve test data to CSV for further analysis in Microsoft Excel
Save time, get answers
- On-site logging of HART data. After logging by the 710 on-site, the ValveTrack software can upload the HART configuration of up to 20 HART devices in your installation and output data in .csv or in .txt format.
- Logged mA loop measurements and HART data from a given transmitter can be logged for the purpose of troubleshooting and fine-tuning the loop. The data logging function has a selectable recording interval from 1 to 60 seconds and a logging capacity of 4910 records or 99 individual sessions. Each record contains the mA measurement and all four process variables.
Main product features
- Test procedures rated Good, Marginal or Poor for control valves
- Generic HART communication
- Highest-in-class accuracy of 0.01% of measurement or source value
- Compact rugged design
- Intuitive user interface with 'Quick-Set' button for quick set-up and user-friendliness
- 24 V DC loop power supply with mA measurement mode (-25% to 125%)
- Resolution of 1 µA in mA ranges and 1 mV in voltage ranges
- Built-in selectable 250Ohm resistor for HART communication
- Simple two-wire connection for all measurements
- Automatic shutdown for longer battery life
- Variable step and rise time in seconds.
HART communication
The Fluke 710 has a built-in HART modem for communication that supports the following HART commands
- Read PV (primary variable) information from sensor
- Read information about PV output
- Read out and write PV unit type, tag ID name, descriptor and message
- Read out and write ranges for the primary variables (upper and lower range values)
- Activating or deactivating fixed current mode
- Setting zero point shift
- Setting zero point DAC (mA output 4 mA)
- Set gain DAC (mA output 20 mA)
HART commands for valves
The 710 provides the following unique HART commands for control valves: Automatic adjustment of the valve controller